Winner2009
Auto Wet Dispersion Laser Particle Size Analyzer

The Winner2009 adopts the Mie scattering principle to deliver reliable and repeatable particle size analysis across a broad measurement range of 0.01μm to 2000μm (or 1200μm). Designed for versatility, it is suitable for a wide variety of applications across multiple industries.
Equipped with a dual-beam and multi-spectral detection system along with side light scattering technology, the Winner2009 significantly enhances measurement precision, sensitivity, and overall performance, representing the leading edge of domestic technology in particle size analysis.
Overview
The Winner2009 Fully Automatic Wet Laser Particle Size Analyzer utilizes the Mie scattering principle to deliver accurate and repeatable particle size measurements over a wide range of 0.01μm to 2000μm (or 1200μm). Designed for diverse application needs, it provides reliable performance across various industries.
Featuring a dual-beam system, multi-spectral detection, and side light scattering technology, the Winner2009 significantly enhances testing precision and sensitivity. It stands as a representation of advanced domestic technology in the field of particle size analysis.
Main Specifications
Model Name | Winner2009A | Winner2009B | ||
Standard | ISO13320-1:2009, GB/T19077.1-2016, Q/JWN001-2024 | |||
Principle | MIE scattering principle | |||
Measuring Range | 0.01μm-2000μm | 0.01μm-1200μm | ||
Channels Number | 97 | 95 (different photo detectors & optical light path) | ||
Accuracy error | <1% (Deviation of D50 on national standard sample) | |||
Repeatability error | <1% (Deviation of D50 on national standard sample ) | |||
Light source | High performance He-Ne Laser (λ= 632.8nm, P>2MW) Lifetime>25000hours | |||
Dispersion Method | Ultrasonic | Frequency:40KHz Power:35W, Time: ≥1S ,Anti-dry function | ||
Stir | Revolutions Speed: 0-3000RPM (Adjustable) | |||
Circulate | Rated Flow:8L/min Rated Power:10W | |||
Sample Pool | Volume:350mL | |||
Micro-Sample Pool | Volume: 10mL ( Available) | |||
Operation Mode | Full automatic/ manual control, freely choose | |||
Resolution | Free distribution truly reflect particle size distribution | |||
Optical bench alignment system | Japan Canon lens, Full automatic,precision is up to 0.1um | |||
Software function | Analysis mode | Free Distribution, R-R Distribution, Logarithm Normal Distribution, Mesh number classification etc. | ||
Statistic Method | Volume Distribution, Quantity Distribution
| |||
Statistic Comparison | Several Testing Results of samples Different batches of samples testing result, Samples before and after processing, Test result of samples in different time.
| |||
User-defined Analysis | Figure out percentage according to the particle size Figure out particle size according to the percentage Figure out percentage according to the particle size range Meet demands of representation of particle test in different industries | |||
Test Report | Word, Excel,Photo( Bmp), Text etc | |||
Multiple-language Support | Multiple language Support | |||
Intelligent operation | Automatically control water inflow, dispersion,test and analysis.Better Repeatability after remove human-factor | |||
Testing speed | <2min/time | |||
Outer dimension | 1200mm*370mm*560mm | |||
Net weight | 70 KG |
Main Features:
Overview
1) Patented Optical Path Design
Winner’s patented optical path design utilizes converging light and Fourier transform techniques to overcome traditional aperture limitations, allowing scattered light to be fully captured. A dual-spectrum design enhances measurement precision across a wide range—from 0.01μm to 2000μm. Additionally, the orthogonal dual-laser setup, aided by a semiconductor auxiliary laser, expands the detection angle from 45° to 135°, ensuring comprehensive signal reception from all directions.
2) Expanded Particle Size Range
The Winner2008A features an extended measurement range down to 0.01μm, making it especially suitable for sub-micron particles. This advanced capability ensures results that meet international standards for fine particle analysis.
3) Fully Automated Built-in Wet Dispersion System
Integrated ultrasonic dispersion, mechanical stirring, and circulation pathways ensure uniform particle dispersion throughout testing. This built-in system avoids common issues such as uneven distribution or particle sedimentation associated with external dispersion setups, thus improving test accuracy.
4) Intelligent Operation Mode
With a smart automatic testing mode, users only need to add the sample and click “Test.” The instrument handles all subsequent processes—water supply, dispersion, circulation, testing, cleaning, data recording, analysis, saving, and printing—automatically. This reduces workload and eliminates human error, enhancing both accuracy and reliability.
5) Automatic Optical Path Alignment
Equipped with four hybrid stepper motors and a micro-precision of 0.1μm, the optical alignment system continuously optimizes the light path. This automation eliminates manual adjustments, ensuring stability and precision throughout testing.
6) Reliable Calibration Method
The system uses certified national standard materials for calibration across the entire measurement range. With simple procedures and a calibration frequency of just once per year, this ensures long-term reliability and ease of maintenance.
7) Proprietary Unconstrained Free-Fitting Technology
Winner’s unique unconstrained free-fitting patent technology enables particle size analysis without being limited by functional assumptions, providing a true reflection of the particle size distribution.
Ⅱ test analysis view
After testing, if necessary, to select records generated an average result, the system analyzes the formation of the recording. When the automatic mode test, without data processing, forming and maintaining a comprehensive system automatically analyzed test records.

Scheme:

Application